Catalog Number 046W1AN00650 |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/23/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3003853072-2019-00166 to 3003853072-2019-00175.
|
|
Event Description
|
It was reported that during the procedure the threads of eight plugs were damaged, the torque handle was out of specification, and the tip of the driver fractured.The tip was removed and the case was completed using an alternate driver and closure tops.There were no patient impacts reported.This is report three of ten for this event.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Udi#: (b)(4).Catalog#: 046w1an00641 final scrdrv shaft rigid ij lot#: a2447214a.Catalog#: 046w0an00002 instinct java blocker lot#: v17069 qty: 8.The event is confirmed with product received.The torque was found to be between 12.44 nm and 14.31 nm with an average of 13.28 nm, which is more than the maximum 11nm.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|