• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SPINE TORQUE LIMITING T-HANDLE; INSTINCT JAVA SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SPINE TORQUE LIMITING T-HANDLE; INSTINCT JAVA SYSTEM Back to Search Results
Catalog Number 046W1AN00650
Device Problem Calibration Problem (2890)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/23/2019
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3003853072-2019-00166 to 3003853072-2019-00175.
 
Event Description
It was reported that during the procedure the threads of eight plugs were damaged, the torque handle was out of specification, and the tip of the driver fractured.The tip was removed and the case was completed using an alternate driver and closure tops.There were no patient impacts reported.This is report three of ten for this event.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Udi#: (b)(4).Catalog#: 046w1an00641 final scrdrv shaft rigid ij lot#: a2447214a.Catalog#: 046w0an00002 instinct java blocker lot#: v17069 qty: 8.The event is confirmed with product received.The torque was found to be between 12.44 nm and 14.31 nm with an average of 13.28 nm, which is more than the maximum 11nm.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORQUE LIMITING T-HANDLE
Type of Device
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
MDR Report Key9510660
MDR Text Key182799154
Report Number3003853072-2019-00168
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
PMA/PMN Number
K111301
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number046W1AN00650
Device Lot NumberA200673402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Age66 YR
-
-