BATTLE CREEK EQUIPMENT THERMOPHORE Back to Search Results
Model Number 155
Device Problems Overheating of Device (1437); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer stated, "pad gets very hot.Too hot and have to shut off, in use i maybe sit on the pad when used on back" customer did not claim injury.Product was returned.The investigator conducted testing for maximum temperatures, possibilities of hot spots, and wattage draw of the device.The investigation supported that the unit was within specification and no defects were observed.Customer admitted to sitting on the product during use.Ifu states, "do not sit on, lie on, or crush pad.Avoid sharp folds".No defects observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE
Type of Device
THERMOPHORE
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
702 south reed rd
fremont IN 46737
Manufacturer Contact
michael green
MDR Report Key9511440
MDR Text Key180777586
Report Number1832415-2019-10947
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number155
Device Lot Number00155305195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2019
Date Manufacturer Received11/27/2019
Type of Device Usage N
Patient Sequence Number1
-
-