• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY QUADRA CRT-D, US MODEL, DF4COMP; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY QUADRA CRT-D, US MODEL, DF4COMP; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3249-40Q
Device Problems Defective Alarm (1014); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient did not feel a vibratory notifier and has presented to clinic with device at eri.It was suggested testing the notifier in clinic under both inductive and rf telemetry.The test was performed and neither the clinician or the patient could feel it vibrating.On (b)(6) 2019, the device was explanted and replaced.The patient was fine.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
New information notes that the device is part of the premature battery depletion advisory.
 
Manufacturer Narrative
The reported field event of vibratory notifier anomaly was not confirmed in the laboratory.The device was at eri upon receipt but had not yet reached eri when explanted.The device was tested on the bench and no anomalies were found.
 
Event Description
Information received also notes that the device did not alert eri on remote website.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY QUADRA CRT-D, US MODEL, DF4COMP
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key9513593
MDR Text Key172618327
Report Number2938836-2019-17614
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734504546
UDI-Public05414734504546
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2013
Device Model NumberCD3249-40Q
Device Catalogue NumberCD3249-40Q
Device Lot Number3713112
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-