On (b)(6) 2019, a carbomedics standard aortic was intended to be implanted in a patient.Upon inspection of the device after the opening of the box, it was identified that the size on the label tag of the device (i.E.23) did not match with the size on the box (i.E.A5-025).Since the serial number of the tag matched the one on the valve box, the surgeon decided to implant the device in the patient assuming that it is a size 25.Despite the labeling discrepancy, no other perceived issues with the device are reported.The procedure was successfully completed using this valve.The patient was not adversely impacted by this event.
|