• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERMO FISHER SCIENTIFIC INC. THERMOFISHER EPREDIA; TISSUE CASSETTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERMO FISHER SCIENTIFIC INC. THERMOFISHER EPREDIA; TISSUE CASSETTE Back to Search Results
Catalog Number B851730YW
Device Problems Loose or Intermittent Connection (1371); Product Quality Problem (1506); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/14/2019
Event Type  malfunction  
Event Description
We have what appears to be a new biopsy cassette in yellow.The lids do not fit properly on these cassettes and we are experiencing loose closure and the possibility of tissue free floating during processing.It appears the newer cassettes are lighter in color than our original ones.Also neither the older lids nor the ones that are supposed to be the companion to the lighter colored cassettes is fitting.I am highly concerned about case integrity and patient safety with this happening.This is a serious issue when considering the integrity of the tissue, case, and providing accurate patient results.(above information was from lab director) these samples are irretrievable.So if we compromise identity, we lose the ability to diagnose.Often there is no additional tissue to go back to.Manufacturer response for tissue cassette, (brand not provided) (per site reporter).Hi all, this is very concerning and i certainly understand the gravity of this situation.I'm copying [] on this message as she represents the manufacturer and needs to be included here.All of the cassettes¿biopsy and slotted in all colors are changing to the new part numbers.This has been taking place gradually over the past few months, with the majority of changes occurring over the past few weeks.Some colors were eliminated (as you know), but otherwise, this transition was to be seamless.I'm listing all of your previous part numbers along with the replacement part numbers below.Have you started using any of these new cassettes in other styles/colors? i'd like to determine if the yellow part number is the only problematic color or if this is more widespread.All of the new part numbers begin with "b851.".(email response from [], thermofisher sales rep).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOFISHER EPREDIA
Type of Device
TISSUE CASSETTE
Manufacturer (Section D)
THERMO FISHER SCIENTIFIC INC.
168 third avenue
waltham MA 02451
MDR Report Key9523341
MDR Text Key172837200
Report Number9523341
Device Sequence Number1
Product Code IDZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/26/2019,12/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberB851730YW
Device Lot NumberI17C9Q1
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/26/2019
Event Location Hospital
Date Report to Manufacturer12/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-