• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX54OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX54OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS Back to Search Results
Model Number 1218-87-354
Device Problem Noise, Audible (3273)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Swelling (2091); Ambulation Difficulties (2544); Test Result (2695)
Event Date 01/01/2017
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).
 
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.No device associated with this report was received for examination.A worldwide complaint database search made on liner lot 2275341 & head lot 2377645 found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Database searches made on the stem lot 2343755 identified one (b)(4) previous complaint.Database searches made on the cup lot bh6j11000 identified one (b)(4) previous complaint.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
 
Event Description
It was reported that the patient have been facing inconvenience pain in the hip area, too much of joint noise and other issues.Also experiencing loud and continuous cracking noise with every movement and leading to limping.This is affecting by daily routine, hampering the patient's work and mental peace.The patient is scared and concerned with the rate of increase in metal levels in blood.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : update 21-jul-2021: the investigation was re-opened upon receipt of the device.Examination of the returned device identified areas of wear and corrosion.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot : the alleged adverse symptoms, and the product code reported, are associated with the articulation between depuy metal-on-metal products.It has been determined that should additional reports be identified for related metal-on-metal complications, a device history record (dhr) review is not required, as the components have been removed from the market.Therefore, a complaint database search and/or device manufacturing review will not be performed for the reported product associated with this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE MTL INS NEUT36IDX54OD
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9526021
MDR Text Key185396049
Report Number1818910-2019-124793
Device Sequence Number1
Product Code OVO
Combination Product (y/n)N
PMA/PMN Number
P090002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1218-87-354
Device Catalogue Number121887354
Device Lot Number2275341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2020
Date Manufacturer Received07/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
-
-