The quality control (qc) that was run before (at 11:53) gave results higher than the range for level 2.The level 2 qc was repeated and gave the result of 24.Quality control results (miu/ml): level 1 (k40981) result: 6.4, level 2 (k40982) result: 40, level 3 (k40983) result: 410.2.Per siemens specifications: level 1 qc range: 3.51 - 11.1, level 2 qc range: 19.1 - 31.1, level 3 qc range: 153 - 225.On 12/16/2019, the system was checked by the siemens customer service engineer (cse).The cse replaced the reagent probes and realigned.Siemens healthcare diagnostics is investigating the cause of the discordant atellica im thcg results.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with the patient's medical history, clinical presentation, and other findings." the ifu states in the limitations section: "all in vitro assays can generate erroneous results, both clinically false positive results (test results suggesting a condition that is absent) and clinically false negative results (test results failing to identify a condition that is present)." mdr 1219913-2019-00281 (initial testing) was filed for the same event.
|
Siemens filed the initial mdr 1219913-2019-00282 on december 27, 2019.April 09, 2020 additional information: the sid (b)(6) was run for thcg and result of 65.6 was generated.Two of the three levels of quality control (qc) were out of range that day.The qc results were as follows: result units: miu/ml: level 1 (k40981) result: 6.4 (3.51 - 11.1).Level 2 (k40982) result: 40 (19.1 - 31.1).Level 3 (k40983) result: 410.2 (153 - 225).The patient sample was repeated on another instrument and a result of 4.8 was generated and considered to be the correct result.The qc level 2 control was also repeated and a result of 24 was generated which is within the range.Review of the ic debug logs showed that the patient sample ran without error.Per the service report, the reagent probes were replaced and realigned.Based on the information given the incorrect result was due to an increase in rlu therefore we would not expect to see any issues with the aspiration traces and replacing and realigning the reagent probes would not correct this issue.There was no definitive root cause found as a result of the investigation.Follow up with the local team indicated that the instrument is operational, and the customer is reporting thcg results.The instrument is performing within specifications.No further evaluation of the device is required.Mdr 1219913-2019-00281 supplemental report 1 was filed for the same event.
|