Model Number DL-344-10-K
Device Problems Break (1069); Impedance Problem (2950)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted lead was returned to neuropace for analysis.Investigation of the returned lead confirmed damage to the distal end coil.Neuropace was unable to identify the root cause for the damage.
 
Event Description
During a routine ecog data review in (b)(6) 2018, a change in the signal and saturation events was noted for the left depth lead.Review of the lead impedance and ecog data was performed and the data was consistent with a possible lead break.There was no report of the patient falling or hitting their head at the time of the noted signal change.(b)(6) 2018, the settings for the left lead were subsequently modified by the treating center.During a routine neurostimulator replacement procedure in (b)(6) 2019, the left depth lead was explanted and a new lead was implanted and connected to the neurostimulator without complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key9527718
MDR Text Key198740936
Report Number3004426659-2019-00063
Device Sequence Number1
Product Code PFN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDL-344-10-K
Device Catalogue Number1007607
Device Lot Number19985-1-1-1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age22 YR
-
-