During a routine ecog data review in (b)(6) 2018, a change in the signal and saturation events was noted for the left depth lead.Review of the lead impedance and ecog data was performed and the data was consistent with a possible lead break.There was no report of the patient falling or hitting their head at the time of the noted signal change.(b)(6) 2018, the settings for the left lead were subsequently modified by the treating center.During a routine neurostimulator replacement procedure in (b)(6) 2019, the left depth lead was explanted and a new lead was implanted and connected to the neurostimulator without complication.
|