Udi: (b)(4).The actual device has been returned and is currently pending evaluation.Once the evaluation has been completed, a supplemental medwatch report will be sent accordingly. if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
It was reported that during an unspecified veterinary surgical procedure, it was observed that the small battery drive device had no power and once removed from surgery they tested it with other batteries and it slowly started working again.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
Corrected data, the date the device was returned to the manufacturer was reported as december 24, 2019 in the initial report and has been updated to december 23, 2019.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device failed the sticky speed trigger assessment, the device stopped working after running for a little bit, the motor was damaged, an unknown liquid was found on the electronic control unit and other components had water marks and contaminated grease on them.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|