It was reported that left hip revision surgery was performed.During the revision, the bhr cup and bhr head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr head.Similar complaints have been identified for the bhr cup.This will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.Based on the limited information provided, the reported pain, limp and elevated cocr levels, and the intraoperative findings of metallosis and discolored tissue may be consistent with findings associated with metal debris.The lateral positioning of the acetabular component cannot be excluded as a result of the reported clinical symptoms and short external rotators which were noted intraoperatively to be completely destroyed.Without supporting post-primary and pre-revision radiographic images, lab reports and/or the analysis of the explanted components, the source of the reported findings cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and expected postoperative healing/pain cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|