This event has been recorded by zimmer biomet under (b)(4).Concomitant medical products - concomitant therapy: item number: 00882100600, item number: power supply, elec.Dermatome, serial number: (b)(4).Report source- foreign - (b)(6).The investigation is still in progress.Once the investigation is complete a follow up mdr will be submitted.
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, g4, g7, h2, h3, h4, h6, h10.Product evaluation: product review of the electric dermatome (b)(6) by dover on (b)(6) 2020 revealed that the motor was running erratically, the head was damaged, and the calibration was out at the 0 setting only.Product repair: repair of the device was performed by dover on (b)(6) 2020 which included replacement of the following: motor, head, bearings additional repair included recalibration the device, serial number (b)(6), was then tested and functioned properly.Review of the device history records identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|