• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Catalog Number 00882100100
Device Problem Failure to Cut (2587)
Patient Problem No Patient Involvement (2645)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Concomitant medical products - concomitant therapy: item number: 00882100600, item number: power supply, elec.Dermatome, serial number: (b)(4).Report source- foreign - (b)(6).The investigation is still in progress.Once the investigation is complete a follow up mdr will be submitted.
 
Event Description
It was reported that the unit had abnormal movement.Device was tested before surgery.The event did not occur during surgery, and there was no patient involvement.No adverse events were reported as a result of this malfunction.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, g4, g7, h2, h3, h4, h6, h10.Product evaluation: product review of the electric dermatome (b)(6) by dover on (b)(6) 2020 revealed that the motor was running erratically, the head was damaged, and the calibration was out at the 0 setting only.Product repair: repair of the device was performed by dover on (b)(6) 2020 which included replacement of the following: motor, head, bearings additional repair included recalibration the device, serial number (b)(6), was then tested and functioned properly.Review of the device history records identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key9531332
MDR Text Key197770274
Report Number0001526350-2019-01204
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00882100100
Device Lot Number61632117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2019
Was the Report Sent to FDA? No
Date Manufacturer Received05/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-