AVANOS MEDICAL - IRVINE HOMEPUMP C-SERIES, 270 ML, 5 ML/HR (CONTAINS DEHP); ELASTOMERIC LFR
|
Back to Search Results |
|
Model Number C270050 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 27-dec-2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
|
|
Event Description
|
Fill volume: 241 ml.Flow rate: 5 ml/hr.Procedure: unknown.Cathplace: unknown.A fast flow event occurred.The infusion should have run until about 11:30am but finished at 2:00am, 9-hours early.The device should infused over 48-hours.The patient is stable.
|
|
Manufacturer Narrative
|
The device history record for the reported lot number, 0002970697, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 27-feb-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
Search Alerts/Recalls
|
|
|