• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2722-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2019
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and a physical evaluation of the device could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event Description
A clinical care coordinator reported that a fresenius combi set disconnected during a patient¿s hemodialysis (hd) treatment, resulting in blood loss.The heparin line on the arterial end of the bloodline reportedly disconnected from the heparin pump.It was reported that this occurred approximately halfway through the patient¿s treatment.A small amount of blood was observed leaking from the bloodline due to the disconnection.The patient¿s estimated blood loss (ebl) was reported to be less than 5 ml.The operator immediately clamped off the lines and returned the patient¿s blood.The machine, a fresenius 2008t, did not alarm.There were no visible defects on the combi set, and nothing out of the ordinary was noticed during the prime.The reporter confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed their treatment after being set up with new supplies on the same machine.The combi set was discarded after the incident and therefore was not available to be sent back for a physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key9536391
MDR Text Key189065192
Report Number8030665-2019-01964
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2019
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number03-2722-9
Device Lot Number19JR01164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age MO
Date Manufacturer Received12/12/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-