MAUDE Adverse Event Report: ZIMMER BIOMET, INC. EXPLOR 8X28MM IMPL STEM W/SCR; PROSTHESIS ELBOW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. EXPLOR 8X28MM IMPL STEM W/SCR; PROSTHESIS ELBOW Back to Search Results
Model Number N/A
Device Problems Detachment of Device or Device Component (2907); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Foreign Body Reaction (1868); Failure of Implant (1924); Pain (1994); Tissue Damage (2104)
Event Date 01/14/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Remains implanted.
 
Event Description
It was reported that a patient underwent an explor elbow arthroplasty on an unknown date.Subsequently, the patient is being revised on an unknown date due to the locking screw backing out.
 
Event Description
It was further reported that the patient's initial surgery occurred approximately 10 years ago and the revision occurred about 2 months ago.
 
Manufacturer Narrative
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was further reported that as well as having a revision due to the screw backing out the patient also had pain, tissue damage, and metallosis.
 
Manufacturer Narrative
(b)(4).Updated: b4, b5, g4, h1, h2, h3, h6, h10 reported event was considered confirmed as the x-rays showed loosening of the radial head with dissociated screw as well as a small fracture throughout the radius.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-02767.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPLOR 8X28MM IMPL STEM W/SCR
Type of Device
PROSTHESIS ELBOW
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9538492
MDR Text Key173541926
Report Number0001825034-2019-05821
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
PMA/PMN Number
K051385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2015
Device Model NumberN/A
Device Catalogue Number11-210063
Device Lot Number543210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN HEAD; UNKNOWN STEM
Patient Outcome(s) Hospitalization; Required Intervention;
-
-