Model Number N/A |
Device Problems
Detachment of Device or Device Component (2907); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problems
Foreign Body Reaction (1868); Failure of Implant (1924); Pain (1994); Tissue Damage (2104)
|
Event Date 01/14/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Remains implanted.
|
|
Event Description
|
It was reported that a patient underwent an explor elbow arthroplasty on an unknown date.Subsequently, the patient is being revised on an unknown date due to the locking screw backing out.
|
|
Event Description
|
It was further reported that the patient's initial surgery occurred approximately 10 years ago and the revision occurred about 2 months ago.
|
|
Manufacturer Narrative
|
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was further reported that as well as having a revision due to the screw backing out the patient also had pain, tissue damage, and metallosis.
|
|
Manufacturer Narrative
|
(b)(4).Updated: b4, b5, g4, h1, h2, h3, h6, h10 reported event was considered confirmed as the x-rays showed loosening of the radial head with dissociated screw as well as a small fracture throughout the radius.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-02767.
|
|
Event Description
|
No further event information available at the time of this report.
|
|