Model Number MSERIES BI-PHASIC |
Device Problem
Output below Specifications (3004)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device's pacer output was out of specification.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.Evaluation results: the customer's report was observed during initial testing.However, after the device was disassembled to isolate root cause the reported problem could no longer be duplicated.The battery interconnect board was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|