• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY VDW GMBH VDW SILVER RECIPROCATING MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY VDW GMBH VDW SILVER RECIPROCATING MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041163000000
Device Problems Mechanical Problem (1384); Failure to Auto Stop (2938)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a silver reciproc motor does not stop when the value is reached.The event outcome is unknown as of this mdr evaluation.
 
Manufacturer Narrative
Battery defective, replaced, motor smr 1456553 ( includes oil and liquid) defective, replaced with smr 1980907.Power supply, foot control and contra angle were not delivered.Function test without errors.Multiple unsuccessful attempts were made to obtain the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW SILVER RECIPROCATING MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
MDR Report Key9551105
MDR Text Key195215972
Report Number9611053-2019-00490
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-