• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problem Gas/Air Leak (2946)
Patient Problems Arrhythmia (1721); ST Segment Elevation (2059)
Event Date 12/12/2019
Event Type  Injury  
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed that at least seven applications were performed with balloon catheter, afapro28 with lot number 28147, on the date of the event.System notice 50012 ¿the refrigerant delivery path was obstructed¿ was triggered at the beginning of the ablation phase in the first application most likely due to residual moisture within cryoablation system.Eventually the moisture was cleared, and the subsequent applications were performed without any recorded anomalies.Clinical issues were encountered during procedure.In conclusion, the adverse events and alleged air ingress could not be assessed via data analysis.There is no indication of relation of adverse event to the performance and malfunction of the product.The physical product was not retuned for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, st elevation occurred, and the patient became bradycardic as the balloon catheter was inserted into the sheath.A small amount of air was aspirated.The patient was paced from the right ventricle and the st segments became normal immediately.The case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9553327
MDR Text Key185936640
Report Number3002648230-2020-00014
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/25/2020
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number28147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-