BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number 87047 |
Device Problem
High impedance (1291)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/23/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported failure was confirmed through investigation analysis.Visual inspection found died body fluid on the handle, main body tubing, and distal end.Dried saline was also noted on the distal end and inside irrigation ports.Electrical tests revealed no shorts and all other electrodes, sensor and thermocouple resistances measured in spec and were typical.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.X-ray inspection revealed no abnormalities.A lumen leak test was performed which the device failed.This indicates a fluid leak inside the shaft.Fluid inside the distal end cavity can cause electrical issues.
|
|
Event Description
|
Reportable based on device analysis completed on december 09 2019.It was reported that high impedance values were observed.During a cavotricuspid isthmus ablation with a intellanav mifi open-irrigated catheter, high impedance values were observed when the catheter was removed from the patient in order to perform a validation map.After the catheter was re-introduced into the patient, high impedances values were still noted.The procedure was completed with another ablation catheter.No patient complications were reported and the patient's current condition is fine.However, returned device analysis revealed a fluid leak inside the shaft.
|
|
Search Alerts/Recalls
|
|
|