• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 DEPUY ASR XL FEM IMP SIZE 45; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 DEPUY ASR XL FEM IMP SIZE 45; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 999890145
Device Problems Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Test Result (2695); No Code Available (3191)
Event Date 10/31/2012
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
Scf and medical record received.Scf alleges pain and alval/soft tissue.After review of medical record, operative note reported loosening of the acetabular and femoral component.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected:h6(device),h6 product complaint # (b)(4).Investigation summary - the asr platform was voluntarily recalled from the market in (b)(6) 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd (b)(4).Ongoing post market surveillance is conducted per our procedures for this product.Device history lot - null.Device history batch - null.Device history review - null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPUY ASR XL FEM IMP SIZE 45
Type of Device
ASR HIP SYSTEM : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key9558958
MDR Text Key185356499
Report Number1818910-2020-00906
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number999890145
Device Lot Number2721770
Was Device Available for Evaluation? No
Date Manufacturer Received02/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age44 YR
-
-