The reported event could not be confirmed due to the condition of the sample.Visual inspection noted one male external catheter was received.Visual evaluation noted the device was used and cut into pieces on return.The results of this investigation were inconclusive due to poor sample condition.A potential root cause for this failure could be mechanical failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|