• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL CEM STEM HO S9; CORAIL AMT CEMENTED IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL CEM STEM HO S9; CORAIL AMT CEMENTED IMPLANTS : HIP FEMORAL STEM Back to Search Results
Catalog Number L96509
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Intraoperatively during opening of the inner sterile package, a hole was noticed in the packaging material.Another implant was available and used without any issue.No surgical delay and no patient harm.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: following investigation, it was concluded that it was likely that the packaging damage occurred during the manufacturing process.Capa-008997 has been raised with regard to this manufacturing issue the complaint shall be entered onto the complaints database and monitored through trend analysis.Should further information be received, then the complaint shall be reviewed and further investigation completed as required.
 
Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL CEM STEM HO S9
Type of Device
CORAIL AMT CEMENTED IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key9560219
MDR Text Key188801711
Report Number1818910-2020-00943
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup,Followup,Followup
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberL96509
Device Lot NumberD18022537
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Date Manufacturer Received03/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-