Device Problem Patient Device Interaction Problem (4001)
Patient Problem Irritation (1941)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
Event description: an event was reported where 4 mib plates were removed due to "hardware irritation".This event was initially collected through trilliant's annual customer feedback survey and specific case details have not been received at this time for any event.This ccr is being closed and four (4) mdr's are being submitted on or before 01/08/2020 in accordance with timeliness and applicable reporting requirements in current versions of sop (b)(4), fda medical device reporting, and sop (b)(4), customer complaint reports.This is report 4 of 4.The below sections will be considered, and if more information becomes available this investigation will be supplemented.Review of surgical technique: a review of the surgical technique cannot be completed at this time.Dhr review: the lot numbers for the mib plates cannot be accurately determined at this time as this surgeon regularly operates at different facilities with different mib system trays and very limited information is available at this time.If and when additional details become available, a dhr review will be completed if lot numbers become known.Visual / dimensional inspection: parts have not been returned at this time and a visual or dimensional inspection cannot be completed.Simulated use testing: due to the nature of the complaint, it is unlikely that simulated use testing can be completed for this event.Evaluation of similar complaints: one other event, ccr 19-04-009, reported that a tiger cannulated screw had backed out of a surgical site from an mib case, but no case details or indication that a removal of hardware occurred was provided.The root cause of ccr 19-04-009 was unknown and was reported to trilliant on 04/18/2019.No additional complaints have been received from this surgeon at this time.Summary / root cause analysis: the root cause of these four (4) reported mib hardware removals remains unknown at this time.
 
Event Description
Senior marketing manager (b)(4) received feedback from dr.(b)(6) on 12/09/2019 at the closure of trilliant's 2019 products and services survey.When prompted to provide more details if "poor" or "fair" was selected for any of trilliant's product offerings dr.(b)(6), on (b)(6) 2019 wrote the following comment regarding trilliant's minimally invasive bunion system: "had to remove 4 out 5 mib plates that i placed.Unacceptable rate of hardware irritation." to gain further clarification trilliant sales representative (b)(4) was reached out to by sales support.(b)(4) originally had a previous complaint from dr.(b)(6) in regards trilliant's minimally invasive bunion plating system from a conversation at midwest podiatry conference (reference ccr 19-04-009.) however, the nature of the survey feedback and the complaint seem independent.(b)(4), trilliant sales support representative, followed up with dr.(b)(6) and requested information for all 4 of the removals.As of 01/07/2020, information has not been received for any of the four reported removals.This mdr represents report 4 of 4 associated with ccr 20-01-003.Further follow up will occur about ancillary details about original implantation date and explantation date for each of the four patients with (b)(4) and dr.(b)(6).This mdr is being submitted on 01/07/2020 to document the awareness of an event within applicable regulatory timeframes.Follow up mdr's will be submitted within applicable timelines if and when additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINIMALLY INVASIVE BUNION PLATING SYSTEM
Type of Device
MINIMALLY INVASIVE BUNION (MIB) PLATE
Manufacturer (Section D)
TRILLIANT SURGICAL
727 n shepherd dr
ste 100
houston TX 77007
Manufacturer (Section G)
TRILLIANT SURGICAL
727 n shepherd dr
ste 100
houston TX 77007
Manufacturer Contact
kaylee ludden
727 n shepherd dr
ste 100
houston, TX 77007
8004952919
MDR Report Key9563831
MDR Text Key220978895
Report Number3007420745-2020-00005
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
PMA/PMN Number
K172178
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date12/09/2019
Date Manufacturer Received12/09/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-