ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problem Fitting Problem (2183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/13/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #t93w6p.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled and upon disassembly, the advancer was confirmed to be bent and one clip was found inside clip track.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.The reported complaint was confirmed.
 
Event Description
It was reported that during a laparoscopic gastrectomy, the clip was not fed into the jaws properly during use.It was also reported that the firing force was higher than expected at the firing, and all clips were deployed though the indicator did not function.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
6107428552
MDR Report Key9571658
MDR Text Key202545733
Report Number3005075853-2020-00179
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-