• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TAPERLOC 133 T1 PPS SO 16X152MM T1; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TAPERLOC 133 T1 PPS SO 16X152MM T1; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-103160
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 12/27/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Udi#:(b)(4).Concomitant medical products: tprlc 133 t1 pps so 14x148mm t1, cat# 51-103140, lot# 3892150.Tprlc xr mp t1 pps 13x111mm 111mm t1, cat# 51-145130, lot# 3894668.Tprlc 133 mp type1 pps so 17.0 1, cat# 51-106170, lot# 6037496.Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-00169, 0001825034-2020-00171, 0001825034-2020-00172.
 
Event Description
It was reported debris was identified within the sterile packaging.There was no patient involvement as this was discovered at the distributor's office.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Evaluation of the returned product/photographs provided confirmed there is debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.The reported event is confirmed and sterility of the product is intact.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage causing the foam packaging to become abraded and shed.The event is being addressed through the capa process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERLOC 133 T1 PPS SO 16X152MM T1
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9574261
MDR Text Key201032234
Report Number0001825034-2020-00170
Device Sequence Number1
Product Code MAY
Combination Product (y/n)N
PMA/PMN Number
K101086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number51-103160
Device Lot Number3886526
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-