Model Number MLCT5D20 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
Hematoma (1884)
|
Event Date 12/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently waiting for the device to be returned for evaluation.
|
|
Event Description
|
During power infusion, leakage in the subcutaneous tissue was noted.When the catheter was removed, a small crack was noted around 2cm from the hub.
|
|
Manufacturer Narrative
|
The 5f midline was returned for evaluation.Visual inspection of the device revealed a hole in the lumen approximately 0.3cm in length at the 4cm mark.The lumen also appears to "curve" at the 4cm mark.Medcomp engineering reviewed the device and confirmed that the lumen appears to have burst at the 4cm depth mark as a result of excessive pressure.The lumen stretched and burst.The manufacture records were reviewed and no non-conformances or variances were noted.Root cause cannot be determined.
|
|
Search Alerts/Recalls
|