THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 201-10002
Device Problems Loose or Intermittent Connection (1371); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 12/23/2019
Event Type  Injury  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
It was reported that the centrimag motor had a loose screw and alarm provided that the pump was not connecting properly.The account replaced the motor with the associated backup system.It was reported that no patient was involved with the event.No further information was provided.
 
Manufacturer Narrative
Section g1: correction.Section g2: correction.
 
Manufacturer Narrative
Manufacturer investigation conclusion: the reported event of a loose set screw was confirmed.The centrimag motor (serial #: (b)(6)) was returned for analysis to the european distribution center (edc) and was evaluated and tested.The system was able to function as intended.The locking ring was replaced.Preventative maintenance was performed.A safety test was performed.All tests were performed per procedure.The motor was returned to the customer.The root cause for the reported event was not conclusively determined through this analysis.The device history records were reviewed for the centrimag motor (serial #: (b)(6)) and the motor passed all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual has an emergency/troubleshooting section for the 2nd generation console.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.The 2nd generation centrimag system operating manual section 12.1 entitled "appendix i ¿ primary console alarms and alerts" contains a list of console alarms and alerts, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, OUS
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key9597990
MDR Text Key176979131
Report Number2916596-2019-06149
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number201-10002
Device Lot Number7380918
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2020
Was the Report Sent to FDA? No
Date Manufacturer Received12/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-