(b)(4).Batch #: unk.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was analyzed, and it was determined that it was connected in more than one generator.The device is intended and labeled for single patient use.If the device is connected to multiple generators an alert screen will be displayed indicating ¿adaptive tissue technology features are not available in this device¿.The device was disassembled to inspect the internal components and no anomalies were found.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This device is packaged and sterilized for single use only.Multiple patient use may compromise the device integrity or create a risk of contamination that, in turn, may result in patient injury or illness.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result include, ¿tighten assembly,¿ ¿blade error detected¿ or "relaxed pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.
|