Vyaire complaint #: (b)(4).Device evaluation: d10, g4, g7, h2, h3, h6, h10 results of investigation: the vyaire failure analysis lab received the ventilator and performed a failure investigation.It was determined that the alarm pcba quad opamp was not reading properly causing a higher dela p.Leaks were found and were isolated to the pneumatic regulator luer fitting threads being stripped during assembly.The ventilator was sent to the vyaire factory service department where it was repaired and preventative maintenance was performed.After being repaired the ventilator passed all testing and met all vyaire manufacturer specifications.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|