• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION KIT Back to Search Results
Model Number IPN046311
Device Problem Failure to Advance (2524)
Patient Problem No Information (3190)
Event Date 01/13/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the plunger of syringes from the lots 71f19h2034, 71f18c1795 and 71f19j2203 is defective.Same issue as previously reported october 09th 2019, no loss of resistance.
 
Manufacturer Narrative
Qn#(b)(4).A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per eco-051699 (released 03-dec-2018), supplier ((b)(4)) made the following changes: kz-05501-002 luer plastic lor syringe: a.Changed plunger material from profax 535 to profax 531; b.Changed to new plunger tool; c.Changed to new mold for blue stopper; d.Changed molding location for the plunger and the blue stopper as follows.- plunger: from fleimaplastic in germany to gpe, germany; - blue stopper: from et, germany to psilkon, germany.These effective changes did impact product design and material.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.
 
Event Description
It was reported that the plunger of syringes from the lots 71f19h2034, 71f18c1795 and 71f19j2203 is defective.Same issue as previously reported october 09th 2019, no loss of resistance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9616505
MDR Text Key180317226
Report Number3006425876-2020-00095
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/12/2021
Device Model NumberIPN046311
Device Catalogue NumberJC-05400-B
Device Lot Number71F19H2034
Was Device Available for Evaluation? No
Date Manufacturer Received02/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-