Section h10: (h3) per capa2019-52, the fractured k-wire reported in case (b)(4) was likely the result of unique misuse conditions, resulting in excessive bending force application to the k-wire, that were not identified during validation and verification activities.These may be surgeon-specific and could not be replicated with subsequent simulated use testing.A recall was initiated for the 3.2mm threaded k-wire (321-52-06).(h7) recall.(h9) if action reported to fda under 21 usc 360i(f), list correction/removal reporting number: z-0168-2020.
|