• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL AUTO ENDO5 ML; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL AUTO ENDO5 ML; CLIP, IMPLANTABLE Back to Search Results
Catalog Number AE05ML
Device Problems Leak/Splash (1354); Difficult to Open or Close (2921)
Patient Problem No Information (3190)
Event Date 12/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history review for the product auto endo5 ml lot #73e1900681 investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that clips overlaid during surgery do not develop as are result of biting.
 
Event Description
It was reported that clips overlaid during surgery do not develop as are result of biting.
 
Manufacturer Narrative
Qn#(b)(4).Per dhr the product auto endo5 ml lot # 73e1900681 was manufactured on 05/27/2019 a total of (b)(4) pieces.Lot was released on 06/03/2019.Dhr investigation did not show issues related to complaint.The customer returned one unit ae05ml autoend05 ml for investigation.The returned sample was visually examined with and without mag nification.Visual examination of the returned device revealed that the sample appears used as there is biological material present on the device.Reference file (b)(4) for investigation photos.The ifu for this product, l06072, was reviewed as a part of this complaint investigation.The ifu for this product states, "mishandling of appliers may result in improper load and/or closure of the ligating clip." the reported complaint of "clip not opened" was confirmed based upon the sample received.One sample was returned.Upon functional inspection, the first two clips fired properly.However, on the third attempt, no clip loaded but a clip shot out of the channel after the trigger cycle was completed.The next clip was out of position in the channel.The sample was disassembled and it was found that the clips were out of position and stacking on one another.A broken clip was found in the channel.The clip stacking prevented the clips from loading properly into the jaws.Although the reported complaint issue was confirmed based on functional testing, it could not be determined exactly how or when the clips became out of position.A capa has been opened to further investigate this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO ENDO5 ML
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key9622435
MDR Text Key176965716
Report Number3003898360-2020-00121
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
PMA/PMN Number
K152081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/26/2022
Device Catalogue NumberAE05ML
Device Lot Number73E1900681
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2020
Date Manufacturer Received02/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-