• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 3-0 NDLS; FASTNER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 3-0 NDLS; FASTNER, FIXATION Back to Search Results
Model Number N/A
Device Problem Difficult to Advance (2920)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/13/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported the juggerknot mini¿s transparent sleeve cannot slide back during the implant insertion during a hand procedure.The transparent sleeve was stuck.Therefore, the implant could not be used.This occurred with two different implants.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
Visual evaluation of the returned product identified no visible damages to both the suture and step drill.The sleeve was not retracted all the way to the handle and some scuffing was noted.Function check was conforming and the device works as intended.Device history record was reviewed and no discrepancies were found.Review of the product determined that no failure was found as the product functions as intended.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.0MM MINI 3-0 NDLS
Type of Device
FASTNER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9637238
MDR Text Key199990211
Report Number0001825034-2020-00272
Device Sequence Number1
Product Code DZL
UDI-Device Identifier00880304523418
UDI-Public(01)00880304523418
Combination Product (y/n)N
PMA/PMN Number
K140908
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number912082
Device Lot Number353060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2020
Was the Report Sent to FDA? No
Date Manufacturer Received06/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-