Model Number N/A |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.
|
|
Event Description
|
It was reported the juggerknot mini¿s transparent sleeve cannot slide back during the implant insertion during a hand procedure.The transparent sleeve was stuck.Therefore, the implant could not be used.This occurred with two different implants.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
Visual evaluation of the returned product identified no visible damages to both the suture and step drill.The sleeve was not retracted all the way to the handle and some scuffing was noted.Function check was conforming and the device works as intended.Device history record was reviewed and no discrepancies were found.Review of the product determined that no failure was found as the product functions as intended.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|