The returned valve was patent.It met the requirements for leakage, reflux, valve flux, siphon control and preimplantation testing.The valve did not meet the requirements for siphon and pressure flow testing.Biological and crystalline debris were observed inside the delta chamber.The instructions for use cautions, ¿the system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ dhr review performed on lot # e63448 did not indicate any finding correlated to the complaint.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|