On 11/21/2019, neotract was informed of a successfully completed prostatic urethral lift (pul) procedure during which a delivery device did not deploy properly.No patient injury or harm was reported.The device in question was returned to neotract for investigation.On 01/07/2020, neotract's investigation of the device indicated that 32mm of the needle tip was missing.On 01/08/2020, a neotract representative present during the procedure confirmed that the needle fragment was retrieved during the procedure.No additional patient follow-up was required.
|