MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/8 HOLE/195MM/RIGHT; CONDYLAR PLATE FIXATION IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/8 HOLE/195MM/RIGHT; CONDYLAR PLATE FIXATION IMPLANT Back to Search Results
Catalog Number 02.124.408
Device Problem Break (1069)
Patient Problems Bone Fracture(s) (1870); Failure of Implant (1924); Impaired Healing (2378)
Event Type  Injury  
Manufacturer Narrative
Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2018, the patient slipped on the tile at her home and fell to the side.The patient suffered a right femur fracture per imaging.On (b)(6) 2018, the patient underwent orif of a right periprosthetic femur fracture.Postoperatively, 2-1/2 weeks later, the patient complained of burning and swelling around the right knee and towards the right lower extremity.On (b)(6) 2018, the patient felt something break at her previous surgical site.A right distal femur periprosthetic fracture occurred.On (b)(6) 2018, the patient complained of left leg pain after exercise which was found to be fractured hardware on (b)(6) 2019.Due to ailed total knee arthroplasty, she was admitted for a procedure because of pain and the risk to the leg with femoral fracture and loose metal.She underwent revision of right total knee arthroplasty, hardware removal on the right femur on the same day.This report is for one (1) 4.5 mm va-lcp curved condylar plate/ 8 hole/ 195mm/ right.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Concomitant device reported: unk - screws: trauma (part number: unknown; lot number: unknown; quantity: unknown).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: g1 h6 part: 02.124.408, lot: l727874, release to warehouse date: 16 jan 2018, manufacturing site: werk mezzovico, expiration date: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the photo revealed that the va-lcp condylar plate 4.5/5.0 r 8ho l195 had broken at the second combi-hole from the condylar end.No material irregularities could be identified from the returned photos.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the va-lcp condylar plate 4.5/5.0 r 8ho l195.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/8 HOLE/195MM/RIGHT
Type of Device
CONDYLAR PLATE FIXATION IMPLANT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK MEZZOVICO
via cavazz 5
mezzovico 6805
SZ   6805
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key9647023
MDR Text Key182748232
Report Number2939274-2020-00542
Device Sequence Number1
Product Code JDP
UDI-Device Identifier07611819454040
UDI-Public(01)07611819454040
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110354
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number02.124.408
Device Lot NumberL727874
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2018
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - SCREWS: TRAUMA
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-