Model Number A22251C
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2020
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection of the prostate (turp) procedure, the roller ball at the distal end of the hf resection electrode broke off and fell into the patient's bladder.However, no fragment remained inside the patient since it was reportedly retrieved.The intended procedure was successfully completed with a similar device and there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was returned to olympus for evaluation/investigation.The investigation confirmed that the wire at the distal end of the electrode is broken on one side and the roller ball has become detached and is missing.The wire ends and the insulation are melted.This type of damage is caused by excessive force, use-related wear and tear, or improper handling.Therefore, this event/incident was attributed to use error.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key9652392
MDR Text Key198147255
Report Number9610773-2020-00057
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761036815
UDI-Public14042761036815
Combination Product (y/n)N
PMA/PMN Number
K951863
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/04/2024
Device Model NumberA22251C
Device Catalogue NumberA22251C
Device Lot Number1000039176
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
VALLEYLAB FX GENERATOR
-
-