14gx15cm amica probe tip malfunction following standard safety protocol resulting in dislodgement of the probe tip in the liver during ablation.The ablation was aborted within 15 seconds of the first ablation cycle.The probe was removed and a new 14gx15cm amica probe was used to re-attempt ablation.Following standard safety practice the new probe was inserted into the area of the tumor.After 1.5 min into the ablation cycle, the machine abruptly cut off due to similar issues as the first incident.The probe tip was also found to be dislodged in the liver.The lot numbers of these two devices were the same.Lot# 37210.Manufacturer response for apparatus for microwave ablation (amica) probe, amica probe (per site reporter).Https://www.Healthtronics.Com/sites/default/files/resources/amica%20information%20sheet.Pdf the radiology service made the local representative aware of the issue.This product is distributed by endocare, inc.- otherwise, amica doesn't have a united states office.
|