MAUDE Adverse Event Report: G.E MEDICAL SYSTEMS ISRAEL LTD. VIVID I; DIAGNOSTIC ULTRASOUND SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

G.E MEDICAL SYSTEMS ISRAEL LTD. VIVID I; DIAGNOSTIC ULTRASOUND SYSTEM Back to Search Results
Device Problems Fire (1245); Smoking (1585); Sparking (2595); Battery Problem (2885)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2020
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.Udi: not applicable.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Ge healthcare's investigation is ongoing.Legal manufacturer: (b)(4).Device evaluation anticipated, but not yet begun.
 
Event Description
A customer reported that a fire started from a battery-powered ultrasound device, the vivid i.They reported ignition, sparks and smoke from the rear/back of the vivid i.The fire was extinguished by a fire extinguisher (c02 initially and then sprayed water because it could not be extinguished).At the time of the event the vivid i was on a cart and was battery powered (no ac).
 
Manufacturer Narrative
Gehc's investigation has completed.The vivid-i battery was returned to gehc for in-depth analysis, and the investigation concluded that the most probable root cause is related to failure in the lithium ion cell which caused the battery to fail and led to thermal runaway.This event occurs once out of ~1.5 million cells of this model that have been implemented in all ge products.Additionally, there is no process root cause.The damaged device was taken out of service.No further actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVID I
Type of Device
DIAGNOSTIC ULTRASOUND SYSTEM
Manufacturer (Section D)
G.E MEDICAL SYSTEMS ISRAEL LTD.
nativ ha'or street no. 1
haifa 35085 10
IS  3508510
MDR Report Key9671392
MDR Text Key182120731
Report Number9615849-2020-00001
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
PMA/PMN Number
K121062
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-