C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER Back to Search Results
Model Number 165814
Device Problems Deflation Problem (1149); Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the balloon on the foley catheter was not filling at all, and then the fluid was seen coming out of a hole about 1cm up from where the saline flush was attached.
 
Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the balloon on the foley catheter was not filling at all, and then the fluid was seen coming out of a hole about 1cm up from where the saline flush was attached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9671900
MDR Text Key183973539
Report Number1018233-2020-00725
Device Sequence Number1
Product Code GBM
UDI-Device Identifier00801741029745
UDI-Public(01)00801741029745
Combination Product (y/n)N
PMA/PMN Number
K760093
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number165814
Device Catalogue Number165814
Was Device Available for Evaluation? No
Date Manufacturer Received02/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-