Corrected information was provided in d.11.Additional information was provided in d.10., h.3., h.6.And h.10.Evaluation summary: the product was returned.Solution is on the lens.Haptic damage was observed.The file indicated the use of a qualified cartridge.The associated handpiece and viscoelastic were not provided.It is unknown if qualified products were used.The product investigation could not identify a root cause for the reported event.Lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.The manufacturer internal reference number is: (b)(4).
|