ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Device Problem Positioning Failure (1158)
Patient Problem Failure of Implant (1924)
Event Date 10/30/2019
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.Product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during a cataract extraction with intraocular lens (iol) implant procedure, during insertion, the iol spurt out with posterior capsule rent.The iol was exchanged for a different model during the initial procedure.
 
Manufacturer Narrative
Corrected information was provided in d.11.Additional information was provided in d.10., h.3., h.6.And h.10.Evaluation summary: the product was returned.Solution is on the lens.Haptic damage was observed.The file indicated the use of a qualified cartridge.The associated handpiece and viscoelastic were not provided.It is unknown if qualified products were used.The product investigation could not identify a root cause for the reported event.Lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9687680
MDR Text Key186534868
Report Number1119421-2020-00209
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.220
Device Lot Number12522946
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Date Manufacturer Received12/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MONARCH III D CARTRIDGES, 8065977763, HWV
Patient Outcome(s) Other;
-
-