Model Number DM3500 |
Device Problems
Premature Discharge of Battery (1057); Premature Elective Replacement Indicator (1483)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/20/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Further information was requested but not received.
|
|
Event Description
|
It was reported that the patient presented with the implantable cardiac monitor exhibiting premature battery depletion.The elective replacement indicator was triggered 16 month post-implant.No interventions were reported.The patient was stable.
|
|
Event Description
|
New information received notes that the implantable cardiac monitor was explanted and replaced.The patient was stable.
|
|
Manufacturer Narrative
|
The complaint of premature battery depletion was not confirmed.Session record history indicated that device had normal battery depletion.Total projected longevity was 1.72 years and device was implanted for 1.50 years, which exceeded 75% projected longevity and is considered normal.
|
|