Model Number SW790T
Device Problem Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/17/2020
Event Type  Injury  
Manufacturer Narrative
If additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported that there was an issue with s4 set screw.According to the customer complaint."l4/l5 fixation using s4 and sri(spondylolisthesis reduction instrument).When inserting set screws into the screws in l5, the rod was slightly elevated in l4 screws and there was a slight gap between the rod and the screws.In right l4, the surgeon attempted to screw in the set screw using a set screw starter, but the set screw rotated idly, so he removed the set screw in question to replace it with a brand new one.The surgeon used two set screws in the left screw of l4, but both rotated idly in the left side, too.Finally, for the right l4 screw, the surgeon managed to fix the second set screw, and it was successful.For the left, the surgeon attempted carefully to fix one of the two failed set screws, and he succeeded in finally tightening the set screw.There was no need for replacement of pedicle screws.It is unknown exactly how long the surgical procedure was delayed.There was no patient harm.Additional information was not provided nor available / was not available.Additional patient information is not available.The adverse event is filed under (b)(4).Associated medwatch-reports: 9610612-2019-00949 ((b)(4) sw790t).
 
Manufacturer Narrative
Associated medwatch-reports: 9610612-2019-00949 ((b)(4),sw790t).General information: intra operative incident.The screw arrived in decontaminated condition.Consequences for the patient: according to the available information there were no negative consequences for the patient.Investigation: we made a visual inspection of the screw.In the first step we investigated the hexagon.The hexagon exhibit no signs of wear or damages, which were signs of a not correct applied hex- key.In the next step we investigated the bottom of the screw.Here we found the typically circular wear of a not proper tightened screw, respectively a screw who was tightened over a not correct placed rod.A sample of the bottom and the wear marks of a correct inserted and tightened set screw is shown.Batch history review: the manufacturing documents have been checked and found to be according to specification valid during the time of production.There are no further complaints with this lot at hand.Conclusion and root cause: the root cause for the problem is most probably usage related.Rationale: the wear shapes on the bottom of the screw are a sure hint for tightening with a not correct (tilted) applied rod.A material failure or a manufacturing error can be excluded.Corrective action: according to sop sa-de13-m-4-2-04-000-0 (corrective action & preventive action) a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S4 SET SCREW NEW VERSION
Type of Device
SPINE SURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key9702351
MDR Text Key179416244
Report Number9610612-2019-00950
Device Sequence Number1
Product Code MNI
Combination Product (y/n)N
PMA/PMN Number
K032219
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSW790T
Device Catalogue NumberSW790T
Device Lot Number52457096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Date Manufacturer Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-