Device evaluated by manufacturer: the device was returned for analysis.A visual inspection revealed that only accustick device was returned for analysis (cannula, dilator, sheath, introducer needle and stylet.).No visual issues were observed with the introducer needle and stylet.The devices did not have any broken section.Residues were observed over the dilator indicating use and handling of the device.The distal section of the cannula, dilator and sheath had a severe kink.The tip of the dilator was damaged.A functional inspection revealed that the components could not be taken apart due to severe kinking at distal section.
|