• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC LEVEL 1 HOTLINE LOW FLOW SYSTEM; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC LEVEL 1 HOTLINE LOW FLOW SYSTEM; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number CON-HL-90
Device Problem Electrical Power Problem (2925)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported the device loses power after running for a while.No patient injury or complications were reported in relation to this event.
 
Manufacturer Narrative
One fluid warmer was returned for evaluation.Visual inspection of the device found the enclosure dirty, missing all of the rubber feet, line cord is old and worn, switch label is peeling off, and noted to contain old style pcb and drain fitting.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.Device underwent functional testing by filling the tank with water and powering on.This confirmed the reported customer complaint due to a bad pcb board.The problem source however has been determined to be unknown.While no definitive problem source to the reported issue could be determined, this investigation revealed no intrinsic evidence to suggest a cause of issue related to manufacturing.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEM
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC
6000 nathan lane north
minneapolis, mn
MDR Report Key9715177
MDR Text Key179715869
Report Number3012307300-2020-01107
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCON-HL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2020
Date Manufacturer Received03/23/2020
Patient Sequence Number1
-
-