SORIN GROUP ITALIA S.R.L. - CRM FACILITY PLATINIUM; DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER
|
Back to Search Results |
|
Model Number PLATINIUM VR 1240 |
Device Problems
High impedance (1291); Failure to Sense (1559); Inappropriate/Inadequate Shock/Stimulation (1574); No Pacing (3268)
|
Patient Problem
No Information (3190)
|
Event Date 01/20/2020 |
Event Type
Injury
|
Event Description
|
Reportedly on (b)(6) 2020, 2 days after implantation, the patient received an inappropriate shock following lead dislodgement, which caused the icd to interpret an atrial fibrillation as a ventricular fibrillation.The lead was then repositioned and all electrical performances were normal upon leaving the operating room.During the follow-up performed on (b)(6) 2020, lead impedance and coil continuity measurements were above 3000 ohms.No sensing and flat egms were observed.On (b)(6) 2020, a re-intervention was performed and the lead was found properly screwed.Tests with a pacing system analyzer (psa) revealed regular lead performances (threshold was 1 v, 0.5 ms; impedance was 400 ohms and detection was normal).The physician then reconnected the lead to the icd, however the same abnormal behavior was observed: lead impedance and coil continuity measurements were saturated at 3000 ohms, no sensing and flat egms were observed.The subject icd was replaced and returned for analysis.No anomaly was observed with the new icd.Preliminary analysis revealed that, due to a wrong software management under specific conditions, the device entered a blocked state, resulting in absence of pacing and sensing.
|
|
Event Description
|
Reportedly on (b)(6) 2020, 2 days after implantation, the patient received an inappropriate shock following lead dislodgement, which caused the icd to interpret an atrial fibrillation as a ventricular fibrillation.The lead was then repositioned and all electrical performances were normal upon leaving the operating room.During the follow-up performed on (b)(6) 2020, lead impendance and coil continuity measurements were above 3000 ohms.No sensing and flat egms were observed.On (b)(6) 2020, a re-intervention was performed and the lead was found properly screwed.Tests with a pacing system analyzer (psa) revealed regular lead performances (threshold was 1 v, 0.5 ms; impedance was 400 ohms and detection was normal).The physician then reconnected the lead to the icd, however the same abnormal behavior was observed: lead impendance and coil continuity measurements were saturated at 3000 ohms, no sensing and flat egms were observed.The subject icd was replaced and returned for analysis.No anomaly was observed with the new icd.Preliminary analysis revealed that, due to a wrong software management under specific conditions, the device entered a blocked state, resulting in absence of pacing and sensing.
|
|
Event Description
|
Reportedly on (b)(6) 2020, 2 days after implantation, the patient received an inappropriate shock following lead dislodgement, which caused the icd to interpret an atrial fibrillation as a ventricular fibrillation.The lead was then repositioned and all electrical performances were normal upon leaving the operating room.During the follow-up performed on (b)(6) 2020, lead impedance and coil continuity measurements were above 3000 ohms.No sensing and flat egms were observed.(b)(6) 2020, a re-intervention was performed and the lead was found properly screwed.Tests with a pacing system analyzer (psa) revealed regular lead performances (threshold was 1 v, 0.5 ms; impedance was 400 ohms and detection was normal).The physician then reconnected the lead to the icd, however the same abnormal behavior was observed: lead "impedance" and coil continuity measurements were saturated at 3000 ohms, no sensing and flat egms were observed.The subject icd was replaced and returned for analysis.No anomaly was observed with the new icd.Preliminary analysis revealed that, due to a wrong software management under specific conditions, the device entered a blocked state, resulting in absence of pacing and sensing.
|
|
Manufacturer Narrative
|
Returned device analysis confirmed that the device was in a permanent blocked state, resulting in absence of pacing and sensing.
|
|
Search Alerts/Recalls
|
|
|