The hillrom technician found the external buzzer board needed to be replaced.The external buzzer is the source of the brake not set alarm.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in january 2020.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the external buzzer board to resolve the issue.Based on this information, no further action is required.
|