ST. JUDE MEDICAL CATD DRAGONFLY OPTIS KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number C408646
Device Problems Material Twisted/Bent (2981); Positioning Problem (3009)
Patient Problem Vascular Dissection (3160)
Event Date 01/27/2020
Event Type  Injury  
Manufacturer Narrative
One dragonfly was returned to the manufacturer of analysis.The results of the investigation concluded that the catheter had been kinked between the lens and the guidewire exit port.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported positioning issue is consistent with the kink.The cause of the kink is consistent with damage during use.
 
Event Description
During the procedure there was a vessel dissection in the left main into the lad.There was difficulty tracking the tortuous vessel and was continuously buckling back on itself at the rx join.A stent was placed over the dissection, and another catheter was used to confirm the stent position.The patient remained stable throughout the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY OPTIS KIT BOX IMAGING CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
Manufacturer (Section G)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key9732281
MDR Text Key188855745
Report Number3009600098-2020-00003
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K141453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/29/2021
Device Model NumberC408646
Device Catalogue NumberC408646
Device Lot Number7054334
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-