Siemens filed initial mdr 1219913-2020-00072 on (b)(6) 2020.Additional information (b)(6) 2020: customer confirmed that sample plid334 is the same sample as plid3341so00 listed in the raw data attachment.Customer confirmed that a correct result of 4.2 ng/ml was reported to physician for sample pkiz760; no other results were reported to physician for this sample.Customer confirmed that sample plid334 is from a 79 year old male; no other patient information was provided.Customer confirms that becasue the reported result was aligned with expectations, this sample was not tested with other methodologies.Additional information (b)(6) 2020: a discordant high atellica immunoassay (im) alpha fetoprotein (afp) result was obtained for sample plid334 on (b)(6) 2020.The sample was repeated on the same instrument with the same reagent pack and calibration.The 3 repeat results align with the physicians expected results for the sample, while the original result and 1 repeat result were elevated (discordant).A service request was opened on (b)(6) 2020, however no service was performed because a precision study was conducted on (b)(6) 2020 with 10 replicates.The results of the precision study were within the criteria documented in the assay instructions for use.As such, the sample was not requested for analysis by siemens.The probable cause of the discordant results is incomplete wash or aspiration, or base dripping in the luminometer, which can be resolved with routine troubleshooting.Siemens investigation is complete.Based on the investigation, no product problem was identified.Mdr 1219913-2020-00071 supplemental report 1 was filed for the initial discordant result.
|