• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-PFO-035
Device Problems Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2020
Event Type  malfunction  
Manufacturer Narrative
The reported event of a twisted appearance on the pfo at the connection with the delivery cable could not be confirmed.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2020, a 35mm amplatzer pfo occluder (pfo) was selected for implant.During device preparation outside the patient, the physician noticed that the connection part between the delivery cable and the pfo appeared as if twisted.The physician continued with the procedure, but became concerned about the issue again when the pfo was deployed in the defect.The pfo was removed from the patient, and another pfo with the same size was implanted with no adverse patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
pamela yip
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key9733572
MDR Text Key183953710
Report Number2135147-2020-00072
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806011301
UDI-Public00811806011301
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-PFO-035
Device Catalogue Number9-PFO-035
Device Lot Number7162169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-